1.Working principle of Dissolved Oxygen Sensor
Dissolved oxygen sensor is a reliable, high-precision dissolved oxygen sensor for a variety of water and wastewater applications.The fluorescence dissolved oxygen sensor is designed based on the principle of quenching of excited fluorescence by specific substances in physics. When the excitation light shines on the fluorescent substance on the surface of the fluorescent membrane head, the fluorescent substance is excited and emits fluorescence. The extinguishing time of the fluorescence is affected by the concentration of oxygen molecules on the surface of the fluorescent membrane head. The concentration of oxygen molecules can be calculated by detecting the luminous lifetime of the fluorescence, and the final value can be output after temperature, air pressure and salinity compensation.
2. Dissolved oxygen sensor technical parameters
Measuring principle | Fluorescence lifetime method |
Measuring range | 0-20mg/L |
Resolution | 0.01mg/L |
Accuracy | ±2%, ±0.3℃ |
Temperature compensation | Automatic temperature compensation |
output method | RS-485 (Modbus/RTU), 4-20mA |
storage temperature | -5~+65℃ |
working environment | 0-50℃ , <0.3MPa |
Installation | Immersion installation, 3/4NPT mounting thread |
Cable length | 5m |
Power consumption | 0.2W, 12V power supply |
powered by | DC: 12V-24V |
Protection level | IP68 |
Calibration method | Two-point calibration |
3. Dissolved oxygen sensor size/installation
1. Dissolved oxygen sensor dimensions
2. Dissolved oxygen sensor wiring diagram
The cable is a 4-core shielded cable:
Red line - power line
Blue wire—ground wire (GND) Green wire—485B
Yellow line—485A
The cable color may vary slightly due to different production batches. Please refer to the cable silk screen for details.
3.Dissolved oxygen sensor installation
Before powering on, the wiring sequence should be carefully checked to avoid unnecessary losses due to wiring errors, and the wiring points should be waterproofed.
4. Dissolved oxygen sensor maintenance and care
1) Sensor outer surface: Rinse the outer surface of the sensor with clean water. If there is still dirt remaining, wipe it with a damp soft cloth. For some stubborn dirt, you can add some household detergent to the water to clean it.
2) Fluorescent membrane head surface: If there is dirt on the surface of the fluorescent membrane head, please rinse it with clean water or wipe it gently with a soft cloth. When cleaning, pay attention to the strength to avoid scratching the measuring area and affecting the measurement accuracy.
3) If the measured value is too high or too low, clean the fluorescent membrane head and stir the membrane head during measurement to prevent foreign objects from adhering to the membrane head surface.
5. Dissolved oxygen sensor calibration
5.1 Zero point calibration
Add 5g of sodium sulfite to 95mL of water and stir evenly. Place the sensor in the solution. After the value stabilizes for 3 minutes, perform zero calibration according to the communication instructions.
5.2 Slope Calibration
Place the sensor in air-saturated water, wait for about 3 minutes for the value to stabilize, then perform slope calibration, and perform zero calibration according to the communication instructions.
5.3 Notes
Avoid exposing the inner surface of the fluorescent film to sunlight or strong light. Do not touch the fluorescent film with your hands or sharp objects.
Avoid bubbles adhering to the surface of the fluorescent film during use
Fengtu Water Level Monitoring Station is a state-of-the-art surface water monitoring device designed for automatic monitoring stations in the field with water surface fluctuation filtering processing to ensure the accuracy of measurement results. The product utilizes a horn antenna design to effecti...
Chemical plant explosion-proof weather stationModel: FT-FB01Manufacturer: Shandong Fengtu Internet of Things Technology Co., Ltd.Product features: FT-FB01 is a new generation of industrial-grade meteorological observation equipment based on ultrasonic series sensors. It adopts a modular design metho...
The Air Quality Monitoring Station plays a crucial role in environmental supervision.The Air Quality Monitoring Station can monitor the concentrations of various pollutants and meteorological parameters in the air, such as PM2.5, PM10, SO₂, NO₂, CO, O₃, TVOC, etc., in real-time and continuously,...
Duct air velocity measurement is a common operation in industrial production and environmental monitoring. Its main purpose is to monitor the flow rate of wind velocity inside a pipeline to ensure that the fluid (usually gas) inside the pipeline is within a controlled range.In industrial production,...